This could be a few issues. Firstly, if the water is too high before you install the samples into the device the water level will rise when you displace water with the samples and trays. This displaced water will rise...
FAQ
If I replace a part on my SmarTracker, do I need to recalibrate?
Motherboard computer No, unless the backup of configuration and calibrations was not performed. LVDTs Yes, the new LVDTs will need to be calibrated to the system RTD temperature probe Yes, the new RTD will need to be calibrated to the...
I would like to network my SmarTracker so that I may use the “Network Storage Feature,” how do I accomplish this?
You will need the following hardware: Client PC; 16g or 32g SD Card formatted to FAT32 Network Switch; The Netgear GS305 has shown to be a good option Two Cat5 cables Length will vary depending on your physical set up...
I’m concerned my computer is having issues, can I back up the calibrations, so I don’t lose them?
Yes, navigate to the control panel Insert a 32g or smaller USB formatted to fat32 Navigate to the software menu Password is 5555 Choose backup from the menu Choose pen drive Click the play button Click the check box Store...
My SmarTracker keeps having a saving to USB issue. How can I fix this?
This could be a few things. Let’s start with trying to replace the USB (some USBs do go bad and need to be replaced). Ideally, you want to have a 32g or less USB formatted to fat32 for these systems....