The furnace dimensions are 21.75x36.75x25.75. We recommend adding 12in on the left and right sides of the furnace and locating it at least 6in from the wall behind the unit.
FAQ
What if my exhaust is going to be longer than 10ft?
After 10ft, a booster fan may be required. Consult an HVAC technician The fan mounted on the NCAT furnace is rated for 65CFM; any booster fan will need to be rated lower than this rating to avoid negative air pressure...
What type of exhaust should I use?
The exhaust tube can reach up to 270C (518F). We recommend a flexible stainless-steel tubing up to 10ft in length. Minimize bends in the tubing for best effect. You should have no more than one 90-degree angle per 10 feet...
What plug type does the NCAT come with from the factory?
The NCAT does not come with power plug attached, this is left for the local electrician to work with local codes and requirements.
What if my power does not match the model of the furnace?
You will need a “buck-boost” transformer; this can be purchased at HMA Lab Supply; PN NC-2905